Model Number 5638
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A manufacturing record review was completed and zero nonconformances were found therefore, supporting the device met material, assembly and performance specifications.A returned product evaluation confirmed the reported failure that the distal tip separated.The catheter also showed signs of sever torque damage along the entire shaft.The turnpike ifu warns; "never advance, withdraw, or rotate an intravascular device against resistance until the cause of the resistance is determined by fluoroscopy.Movement of the catheter or guidewire against resistance may result in separation of the catheter or guidewire tip, other device damage, or vessel injury.".
 
Event or Problem Description
The physician was trying to cross a heavily calcified lesion, and reports that the tip of the catheter broke off.Planned strategy was to fix lad lesion then do a tavr.Due to this incident patient's tavr (transcatheter aortic valve replacement) procedure was done the same day.During the procedure it was noted the tip had migrated into a septal branch where it was speculated it would not harm the patient.It was noted the tip migrated to a septal branch and no further intervention would be required.Patient was being discharged to home on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURNPIKE LP
Common Device Name
CATHETER
Manufacturer (Section D)
VASCULAR SOLUTIONS, INC
6464 sycamore court north
minneapolis MN 55369
MDR Report Key7832365
Report Number2134812-2018-00061
Device Sequence Number11097262
Product Code DQY
UDI-Device IdentifierM20656380
UDI-PublicM20656380
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K151981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2020
Device Model Number5638
Device Lot Number620555
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2018
Initial Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight74
-
-