A manufacturing record review was completed and zero nonconformances were found therefore, supporting the device met material, assembly and performance specifications.A returned product evaluation confirmed the reported failure that the distal tip separated.The catheter also showed signs of sever torque damage along the entire shaft.The turnpike ifu warns; "never advance, withdraw, or rotate an intravascular device against resistance until the cause of the resistance is determined by fluoroscopy.Movement of the catheter or guidewire against resistance may result in separation of the catheter or guidewire tip, other device damage, or vessel injury.".
|