• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with all of the results obtained.The expected fasting blood glucose test result range is 150 to 225mg/dl.The customer feels well and did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored by customer according to specification in the kitchen.The test strip lot manufacturer's expiration date is 06/30/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).The customer was comparing the right hand to the left hand.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7832459
Report Number1000113657-2018-00909
Device Sequence Number1940486
Product Code NBW
UDI-Device Identifier00052569138557
UDI-Public(01)00052569138557
Combination Product (Y/N)N
Initial Reporter StateAL
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2733
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2018
Initial Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-