MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX30008UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 06/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
During the procedure, an attempt was made to use one resolute onyx rs coronary drug eluting stent to treat a lesion in the rca.It was reported that during catheterization the stent was placed in the rca, the guide and stent dislodged during inflation and the stent dislodged and embolized.No patient injury was reported - intervention was required.
 
Additional Manufacturer Narrative
The lesion was in the ostium of the rca.The guide unexpectedly backed out, pulling the stent out of the ostium with it.The stent subsequently dislodged off the delivery catheter and ended up in the carotid.The patient went to neuro and the stent was snared and removed.The patient is doing fine.The customer was determined the stent was not at fault.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7832589
Report Number9612164-2018-02221
Device Sequence Number1941124
Product Code NIQ
UDI-Device Identifier00643169556966
UDI-Public00643169556966
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2019
Device Catalogue NumberRONYX30008UX
Device Lot Number0008629832
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer08/31/2018
01/24/2019
Initial Report FDA Received Date08/30/2018
Supplement Report FDA Received Date09/27/2018
01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient Weight95
-
-