| Model Number |
PED-500-14 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the distal edge of the pipeline was not opening and appeared trumpeted despite 3 resheathings.As a result the device was removed and the patient successfully treated with another pipeline device.The patient was undergoing surgery to treat an unruptured, saccular, 10mm right paraopthalmic aneurysm through a moderate vessel tortuosity.It was noted that the vessels had a large diameter and that the device was not positioned in a bend.Dual antiplatelet treatment was administered but thepru level was unknown.Post angiographic result showed stasis.More than 50% of the pipeline had been deployed when it failed to open.The devices were prepared as indicated per the ifu.There were no patient symptoms.
|
| |
|