| Catalog Number |
10104006 |
| Medical Device Problem Code |
Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).A field service engineer was dispatched to the customer site.The catalytic converter was replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.(b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported an event of a "smoke" or haze emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.The customer was advised to turn the unit off and leave the room.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue was reviewed from (b)(6) 2018 to (b)(6) 2018 and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." the part was not available for return and further analysis.The assignable cause of the smoke/haze issue is likely due to the catalytic converter.The field service engineer replaced the part and confirmed the sterrad 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue was reviewed from (b)(6) 2018 to (b)(6) 2018 and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." the part was not available for return and further analysis.The assignable cause of the smoke/haze issue is likely due to the catalytic converter.The field service engineer replaced the part and confirmed the sterrad 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.(b)(4).
|
| |
|