Catalog Number 222299
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/04/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement the complaint device was received and evaluated.The anchor was still secured to the inserter and the suture was not in its dedicated place.The anchor appears to be used and there is evidence of tissue debris found on the tip of the anchor.The inserter and sutures were examined and no damage or anomalies were found.The anchor was removed from the inserter with no difficulty.A batch review was conducted and the results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaints for this lot of 293 devices that were released to distribution.The lot review indicates there were no issues with material or assembly.A root cause for the user to have experienced this issue cannot be determined.Based on the overall complaint rate for this failure, at this point in time, no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported that during a rotator cuff procedure, one (1) 5.5 mm healix advance br with orthocord and needles fell apart when the surgeon tried to insert it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE BR W/OC&ND
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7832854
Report Number1221934-2018-54365
Device Sequence Number1171309
Product Code MAI
UDI-Device Identifier10886705021352
UDI-Public10886705021352
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K120078
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/05/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2017
Device Catalogue Number222299
Device Lot Number3764028
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2015
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2014
Initial Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-