• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 8590855
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor applicable imaging films were returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent posterior thoracolumbar fixation due to lumbar spondylolisthesis.Intra-op the screw slipped.The product came in contact with the patient.No patient complications were reported as a result of the event.
 
Additional Manufacturer Narrative
Product analysis: microscopic and visual examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with the misalignment of the mas and set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7833050
Report Number1030489-2018-01187
Device Sequence Number1981412
Product Code KWP
UDI-Device Identifier00721902365263
UDI-Public00721902365263
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K981676
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8590855
Device Lot NumberH5375535
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Initial Date Received by Manufacturer 08/28/2018
Supplement Date Received by Manufacturer09/24/2018
Initial Report FDA Received Date08/30/2018
Supplement Report FDA Received Date10/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-