MAUDE Adverse Event Report: HOLGER ULLRICH ALPHAMAXX SURGICAL TABLE W AUTO DRIVE; TABLE, OPERATING-ROOM, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLGER ULLRICH ALPHAMAXX SURGICAL TABLE W AUTO DRIVE; TABLE, OPERATING-ROOM, AC-POWERED Back to Search Results
Model Number 113322F3
Device Problem Unintended Movement (3026)
Patient Problems Fall (1848); Head Injury (1879); Laceration(s) (1946)
Event Date 08/14/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Investigation is ongoing at this time.Upon completion a follow-up medwatch will be submitted.
 
Event Description
It was reported that at the end of a procedure the head end of the or table moved and the patient slid from the table hitting his head.This resulted in a laceration requiring sutures and a ct scan.(b)(4).
 
Event Description
Manufacturer reference # (b)(4).
 
Manufacturer Narrative
Getinge usa sales, llc (importer) is submitting this report on behalf of the legal manufacturer of the device maquet gmbh, kehler strasse 31, rastatt, germany 76437.Exemption # e2018004.(b)(4), contact person: (b)(6).Initially the customer reported the following malfunction : the end head of the or table moved and the patient slid off the table.A getinge- maquet service tech visited the customer and investigated the device.The described malfunction could not be reproduced.The service tech reported that no one in the room actually saw the table move on its own.After the service tech investigation we assume that the patient was not secured properly on the table.The size of the patient and the angle of the table might have contributed to this event.In the instructions for use (ifu)-the customer is warned that the patient can slip off the table if not secured properly.Quote- ifu " warning! risk of injury! if the patient is not secured during positioning, when adjusting the or table or when positioning the patient (especially when the inclination and /or tilt feature is used), then the patient could slip off the table top in a uncontrolled manner or their legs and legs may hang down.Always secure the patient using suitable aids (e.G.Straps)and maintain continuous observation.And "note if a patient is on the or table: secure the patient prior to moving or adjusting the or table" the service technician inspected the table and found a slight leakage at one of the hydraulic cylinders.This kind of minor leakage could not result in the described malfunction.The leaking cylinder was repaired.Getinge-maquet gmbh provides product failure investigation, analysis and resolution for the device described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALPHAMAXX SURGICAL TABLE W AUTO DRIVE
Type of Device
TABLE, OPERATING-ROOM, AC-POWERED
Manufacturer (Section D)
HOLGER ULLRICH
maquet gmbh
kehler strasse 31,
rastatt 76437
GM  76437
MDR Report Key7833206
MDR Text Key118784976
Report Number8010652-2018-00017
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number113322F3
Device Catalogue Number113322F3
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/21/2019
Event Location Hospital
Date Manufacturer Received08/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-