• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIATRS45AXT
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopy, the stapler was not able to fire.Once closed and the green button activated to lock on the tissue, the reload released wide open.The correct process was tried several times to see if the product would correct itself.Another reload was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual and functional evaluation of the reload found no abnormalities.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition; therefore, a review of the device history record was not performed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopy, the stapler was not able to fire.Once closed and the green button activated to lock on the tissue, the reload released wide open.The correct process was tried several times to see if product would correct itself.Another reload was used to complete the case.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7833937
Report Number1219930-2018-04769
Device Sequence Number11051697
Product Code GDW
UDI-Device Identifier20884521190990
UDI-Public20884521190990
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K133938
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model NumberEGIATRS45AXT
Device Catalogue NumberEGIATRS45AXT
Device Lot NumberN7L1086KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2018
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer11/19/2018
Initial Report FDA Received Date08/30/2018
Supplement Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-