| Medical Device Problem Codes |
Moisture Damage (1405); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (case damage w/moisture) issue.The reporter alleged that the battery compartment was found to be cracked.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or the cartridge compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission: 17-sep-2018.Device evaluation: the device has been returned and evaluated by product analysis on 05-sep-2018 with the following findings: during investigation, moisture damage was observed in the battery compartment.The battery compartment was cracked form the threads to the case seal along the bumper pad, and from the case seal to the bumper pad.
|
| |
|