Unknown taper.Device labeling addresses the reported events as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Ulceration, gastritis, gastroesophageal reflux, heartburn, gas bloat, dysphagia, dehydration, constipation, and weight regain have been reported after gastric restriction procedures.Esophageal distension or dilatation has been infrequently reported.This is most likely a consequence of incorrect band placement, over-restriction or stoma obstruction.It can also be due to excessive vomiting or patient noncompliance, and may be more likely in cases of pre-existing esophageal dysmotility.Deflation of the band is recommended if esophageal dilatation develops.A revision procedure may be necessary to reposition or remove the band if deflation does not resolve the dilatation.Warnings: patients should be advised that the lap-band system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.The manufacturer of the lap-band ap adjustable gastric banding system has designed, tested and manufactured it to be reasonably fit for its intended use.However, the lap-band ap system is not a lifetime product and it may break or fail, in whole or in part, at any time after implantation and notwithstanding the absence of any defect.Causes of partial or complete failure include, without limitation, expected or unexpected bodily reactions to the presence and position of the implanted device, rare or atypical medical complications, component failure and normal wear and tear.In addition, the lap-band ap system may be easily damaged by improper handling or use.
|
|
Taper ii.Supplement #1 - medwatch sent to the fda on 05/oct/2018.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.White particles were noted on the outer surface of the band shell.The band tubing was noted to be separated approximately 0.5 inches past the ss connector.Needle marks were noted on the port septum.Scratches were noted on the port housing and port holes.An air leak test was performed, and no leakage was noted.A fill inspection test was performed, and no blockage was noted when colored di water was passed through the port septum and tubing.Material degradation was observed on approximately 30% of the band shell.The end of the band tubing, separate approximately 12.2 inches past the collar, was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.The end of the band tubing, separated approximately 0.3 inches past the ss connector, was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.
|