Additional information: the customer reported that the device was infusing simple ringer serum and the maintenance protocol was reportedly 10 ml/hour, to maintain deep venous access permeability.Investigation ¿ evaluation: a review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually inspected by quality control and no notable gaps in production or processing controls were noted.The risk specification for this product includes the catheter fracture failure mode and identifies that multiple risk controls are in place to mitigate the risk of this type of failure.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be established.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
|