• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX BR ANCHOR W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX BR ANCHOR W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 222233
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/08/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: visual observation confirms the distal tip of the anchor is broken, but is held in place by suture.The anchor appears to be used and there are evidence of tissue debris found on the anchor and the tip of the shaft.This type of anchor breakage at its distal tip is consistent with historical investigations in which it was determined that the root cause was likely a combination of torque and bending, a user technique issue.Other than this possibility, we cannot discern a root cause for this failure.A batch review was conducted to determine if there were any internal processing issues, which would have contributed to the nature of the product complaint.Our results indicate that this batch was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaints of any type for this lot of (b)(4) devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Event or Problem Description
It was reported by the rep that the surgeon has several years¿ experience using the product, successfully.Before insertion of implant, one (1) 5.5 mm healiz br anchor with orthocord broke inexplicably passing through a typical large cannula always used by surgeon.Strange occurrence.Device will be forwarded on receipt.The procedure was reportedly completed with same like product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX BR ANCHOR W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7835257
Report Number1221934-2018-54416
Device Sequence Number16505300
Product Code MAI
UDI-Device Identifier10886705007790
UDI-Public10886705007790
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Catalogue Number222233
Device Lot Number3748908
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2015
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2015
Initial Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-