| Model Number |
B-2245 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unknown taper.Device labeling addresses the reported event as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Warnings: patients should be advised that the lap-band system is a long-term implant.Explant (removal)and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
|
| |
|
Event or Problem Description
|
|
Reported as: a patient with the lap-band system was reported to have a port leak.The device was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Taper ii.Supplemental #1 - medwatch sent to the fda on 05/oct/2018.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.Black particulate matter was noted on the port housing near the port septum.Needle marks were noted on the port septum.Scratches were noted on the port housing.The port tubing was noted to be kinked near the taper tubing junction.A port leak test was performed, and leakage was noted from one opening on the port tubing at the taper tubing junction.An air leak test was not feasible, as only the access port was received.A fill inspection test was performed, and no blockage was noted when colored di water was passed through the port septum and tubing.Under microscopic analysis, two smooth-edged openings were noted on the port tubing.The openings in the port tubing were located at the taper tubing junction, where the device had been observed to be leaking during the port leak test.The end of the band tubing, separated approximately 0.5 inches past the ss connected, was noted to have striated edges, consistent with damage from a surgical end cut to remove the device.
|
| |
|
Search Alerts/Recalls
|
|
|