Approximate age of device- 10 days.Manufacturer's investigation conclusion: thrombus was confirmed through the evaluation of heartmate ii lvas.Heartmate ii lvas was returned assembled with the driveline (dl) cut approximately 8.5¿ from the pump housing and the distal portion of the dl was not returned.The sealed inflow conduit (inlet tube, flex section, and inlet elbow) was returned attached to the pump¿s inlet port.The sealed outflow conduit (outflow graft, bend relief and outlet elbow) was returned attached to the pump¿s outlet port.The sealed outflow graft bend relief collar was sutured around the outflow graft nut.Upon disassembly of heartmate ii lvas, the outlet elbow revealed a thrombus formation.The thrombus appeared to obstruct majority of the blood flow path.The thrombus formation was adhered to the textured blood-contacting surface of the outlet elbow and its structure suggests that it formed within the outlet elbow over an undetermined amount of time while the pump was supporting the patient.The evaluation could not determine a specific cause for the development of the observed thrombus formation within the outlet elbow.Examination of the remaining blood-contacting surfaces of the returned device revealed no additional depositions or thrombus formations that would have contributed to a functional or flow issue.A correlation between the thrombus formation and the patient¿s respiratory failure could not conclusively be established through this evaluation.Additionally, a specific cause of the reported sepsis and infection could not conclusively be determined through this evaluation.Upon removal of the observed depositions, the device was cleaned.The pump's bearings, rotor, and blood-contacting surfaces were then examined under a microscope and no anomalies were observed.Electrical continuity testing of the driveline revealed no discontinuities or shorts.The pump was reassembled and functionally tested under normal operating conditions using a mock circulatory loop.The data retrieved from that testing revealed normal pump power consumption and pressure values, comparable to what was recorded during the manufacturing process and the device functioned as intended.A review of the device history records revealed the device met applicable specifications.No further information was provided.The manufacturer is closing the file on this event.
|