• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-FEM-CELECT-PT
Device Problems Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Ulcer (2274); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.Name and address for importer site:(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a celect filter on (b)(6) 2015".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2015 via the due to blood clots.Patient is alleging "tilt, vena cava perforation, and organ perforation".Patient further alleges "ulcers, gi bleed, and pain all the time".
 
Manufacturer Narrative
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.Patient code: vessels, perforation of (2135)- listed in ifu, organ(s), perforation of (1987) - listed in ifu, ulcer (2274) not listed in ifu.Device code: unintended movement (3026)- not listed in ifu, appropriate term/code not available (3191) - [device perforation]- listed in ifu.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information provided at this time.
 
Manufacturer Narrative
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).G1) name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.Investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿celect platinum - tilt, vena cava and organ perforation, ulcers, gi bleed, pain".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported pain is directly related to the filter.Unknown if the reported ulcers, gi bleeding is directly related to the filter and unable to identify a corresponding failure mode at this point in time.No relevant notes found on work order or device lot.No other complaints on lot.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key7836299
MDR Text Key118870863
Report Number3002808486-2018-00990
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/24/2018
Device Catalogue NumberIGTCFS-65-1-FEM-CELECT-PT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Distributor Facility Aware Date08/16/2018
Date Manufacturer Received12/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight72
-
-