| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a power (battery life) issue; the battery life reportedly lasts no longer than three days.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission (b)(4) 2018 device evaluation: the device has been returned and evaluated by product analysis on (b)(4) 2018 with the following findings: a review of the black box showed evidence of short battery life issues with drastic voltage drops.Unrelated to the original complaint, the battery compartment was cracked below the grip pad.The returned battery cap was undamaged and able to fit.The ¿ez-prime¿ steps were performed correctly and all current readings were above specifications.The reported ¿short battery life¿ was duplicated.The pump cover was removed and moisture corrosion was found to the continuous glucose monitoring module.
|
| |
|