| Model Number |
980 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the service engineer (se) inspected the device, reviewed the logs and verified the reported issue.The se replaced the pneumatic interface printed circuit board assembly (pcba).The ventilator passed all testing per manufacturing specification and was placed back into clinical use.If the replaced part is returned for failure investigation, a supplement medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a 980 ventilator had a "system failure" error code.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the pneumatic interface printed circuit board was returned for failure investigation.The part was visually inspected and functionally tested with no anomalies observed.Conclusion: there was no fault found.There was no malfunction or product deficiency identified.If information is provided in the future, a supplemental report will be issued.
|
| |
|