• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 6DCFS
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation: health professional (hpro).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, the inner cannula extended well past the tracheostomy tube cannula.It was reported that there was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.A visual and functional test was performed.No reported event was observed since met with the product specifications and no corrective actions are required.Stating information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez,ci 32590
MX  32590
MDR Report Key7836911
Report Number2936999-2018-00556
Device Sequence Number16528110
Product Code JOH
UDI-Device Identifier10884522006620
UDI-Public10884522006620
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2020
Device Model Number6DCFS
Device Catalogue Number6DCFS
Device Lot Number15A0900JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer10/26/2018
11/21/2018
Initial Report FDA Received Date08/31/2018
Supplement Report FDA Received Date11/21/2018
12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-