| Catalog Number |
413404 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer in the united states reported a false susceptible ampicillin/sulbactam result for an escherichia coli esbl positive api survey strain, in association with the vitek® 2 ast-gn73 test kit (lot 5930713203).The customer twice obtained a susceptible result (mic=8) for ampicillin/sulbactam.The expected result for this survey strain was intermediate/resistant.A biomérieux internal investigation will be initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer no longer has their api survey strain for this investigation.Submittal of the isolate is required in order to confirm a vitek 2 discrepancy compared to the reference method.Vitek 2 ast-gn73 lot# 5930713203 met final qc release criteria and passed qc performance testing.
|
| |
|