B. BRAUN MEDICAL INC. DESIGN OPTIONS®; SPINAL EPIDURAL ANESTHESIA Back to Search Results
Catalog Number 530048
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: customer reports that the clinician was inserting catheter through tuohy needle and met resistance when trying to remove the needle.When removing the catheter through the needle, the catheter sheared.The clinician was able to remove the wire portion of catheter at that time.About six cm of catheter remained in patient.A new spinal one level down was done and after spinal was able to remove the remaining portion of the catheter from patient.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Multiple unsuccessful attempts were made to obtain a sample.Without the actual device, a thorough investigation could not be performed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If a sample and/or any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DESIGN OPTIONS®
Common Device Name
SPINAL EPIDURAL ANESTHESIA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key7837121
Report Number2523676-2018-00065
Device Sequence Number1573604
Product Code OFT
UDI-Device Identifier04046964305313
UDI-Public04046964305313
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Catalogue Number530048
Device Lot Number0061605303
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/07/2018
Supplement Date Received by Manufacturer08/07/2018
Initial Report FDA Received Date08/31/2018
Supplement Report FDA Received Date09/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-