DEPUY ORTHOPAEDICS, INC. 1818910 TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-020
Medical Device Problem Codes Corroded (1131); Degraded (1153); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Host-Tissue Reaction (1297); Edema (1820); Unspecified Infection (1930); Inflammation (1932); Memory Loss/Impairment (1958); Pain (1994); Swelling (2091); Injury (2348); Limited Mobility Of The Implanted Joint (2671); Test Result (2695)
Date of Event 08/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Concomitant medical products: asr uni femoral impl size 41; asr tap slv adap 12/14 -1; asr 300 spiked cup size 46; tri-lock bps sz 2 std offset.
 
Event or Problem Description
The patient was revised due to increased metal ion levels.Depuy trilock with tri spike asr shell revision.Doi: unknown; dor: (b)(6) 2018; right hip.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
After review of medical records, the patient was revised due to failed right hip mom replacement resulting to elevated metal ions, pain and memory delay.Operative note reported synovial fluid appeared grayish in color and significant amount of corrosion on the head neck junction.However, there is no lab result provided for elevated metal ions.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
Litigation records received.In addition to what was previously reported.Litigation alleges swelling, inflammation, infection, damage to surrounding bone and tissue, walking difficulty, lack of mobility, metallosis and injury.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is to capture edema and device revision or replacement.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 4 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7837327
Report Number1818910-2018-68238
Device Sequence Number2434532
Product Code KWA
UDI-Device Identifier10603295000730
UDI-Public10603295000730
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K073570
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1012-04-020
Device Catalogue Number101204040
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/06/2018
Supplement Date Received by Manufacturer09/28/2018
01/30/2019
05/21/2019
04/08/2020
04/24/2020
04/24/2020
Initial Report FDA Received Date08/31/2018
Supplement Report FDA Received Date10/01/2018
02/22/2019
05/24/2019
04/20/2020
04/28/2020
04/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-