| Model Number |
G47614 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: non-healthcare professional.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use (ifu) instructs the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide.The ifu instructs the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat® 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat® 2 calibrated tip wire guide.The outside of wire was peeling off (subject of this report).The coating of the wire was peeling off during use and this is not the first one (see related mdr 1037905-2018-00384).According to customer the coating did not peel off because of contact with the elevator of the endoscope.The following clarification was received on 08-aug-2018: they say they have had more wires where the coating of the wire has been peeling off that they just have not put aside (see related mdr 1037905-2018-00384).The wire guide damage was at the patient end, close to the tip, and i am not able to provide more details (subject of this report).A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation; our evaluation of the product said to be involved confirmed the report.Approximately 3.8 cm to 4.6 cm from the distal end, the wire guide covering has accordioned.Approximately 4.6 cm to 5.7 cm from the distal end is a section of bare core wire.No rough surfaces or kinks were found along the wire guide.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use (ifu) instructs the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide.The ifu instructs the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat® 2 calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat® 2 calibrated tip wire guide.The outside of wire was peeling off (subject of this report).The coating of the wire was peeling off during use and this is not the first one (see related mdr 1037905-2018-00384).According to customer the coating did not peel off because of contact with the elevator of the endoscope.The following clarification was received on 08-aug-2018: they say they have had more wires where the coating of the wire has been peeling off that they just have not put aside (see related mdr 1037905-2018-00384).The wire guide damage was at the patient end, close to the tip, and i am not able to provide more details (subject of this report).A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|