Model Number CD3249-40Q
Device Problems Application Program Version or Upgrade Problem (2881); Inappropriate or Unexpected Reset (2959); Operating System Version or Upgrade Problem (2997)
Patient Problems Muscle Stimulation (1412); Undesired Nerve Stimulation (1980)
Event Date 08/16/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in clinic for follow-up.A firmware upgrade was attempted for the implantable cardiac defibrillator.During the upgrade the device went into backup mode.The patient had phrenic nerve stimulation due to the high pacing outputs during the backup mode.The device was successfully restored and the upgrade was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY QUADRA CRT-D, US MODEL, DF4COMP
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7838431
MDR Text Key119273155
Report Number2017865-2018-13083
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734504546
UDI-Public05414734504546
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2013
Device Model NumberCD3249-40Q
Device Catalogue NumberCD3249-40Q
Device Lot Number3718373
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/05/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-