| Model Number |
DELTA |
| Medical Device Problem Code |
Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Data review determined there was no malfunction of the instrument or barcode reader.The cause of the patient number ending changing from "y" to "z" was that the barcode reader was set to a (b)(4) keyboard layout.The barcode scanner was reconfigured to a (b)(4) keyboard layout, which resolved the issue.There was no patient impact, although there is potential for reporting the wrong results to a different patient therefore this incident report was submitted.No additional remedial action required.
|
| |
|
Event or Problem Description
|
|
Customer reported an issue with their rotem delta.They scanned the barcode of a patient's id number which had a letter y at the end (eg.S1234567y) but the last letter for the id number that showed on the rotem delta was z (eg.S1234567z).There was no patient impact.
|
| |
|