• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL NANO 1MM X 2CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035431020
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is 2 of 2 reports.The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection of the device revealed that the delivery wire was kinked, likely due to handling.The main coil was found to be kinked and appears to be physically detached at the main coil junction.During functional testing, coil could not be advanced through the introducer sheath as the main coil was stuck in the tip of the introducer sheath.The physician may have perceived the observed damages (coil delivery wire kink) as break and reported it as such; however, no break was identified on the device.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Event or Problem Description
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection of the device revealed that the delivery wire was kinked, likely due to handling.The main coil was found to be kinked and appears to be physically detached at the main coil junction.During functional testing, coil could not be advanced through the introducer sheath as the main coil was stuck in the tip of the introducer sheath.The physician may have perceived the observed damages (coil delivery wire kink) as break and reported it as such; however, no break was identified on the device.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Event or Problem Description
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL NANO 1MM X 2CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7838732
Report Number3008881809-2018-00377
Device Sequence Number11346906
Product Code HCG
UDI-Device Identifier04546540698001
UDI-Public04546540698001
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2021
Device Catalogue NumberM0035431020
Device Lot Number20388291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer10/05/2018
Initial Report FDA Received Date08/31/2018
Supplement Report FDA Received Date10/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR MICROCATHETER(STRYKER)
Patient Age56 YR
Patient Weight58
-
-