| Catalog Number |
M0035431020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is 2 of 2 reports.The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection of the device revealed that the delivery wire was kinked, likely due to handling.The main coil was found to be kinked and appears to be physically detached at the main coil junction.During functional testing, coil could not be advanced through the introducer sheath as the main coil was stuck in the tip of the introducer sheath.The physician may have perceived the observed damages (coil delivery wire kink) as break and reported it as such; however, no break was identified on the device.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
|
| |
|
Event or Problem Description
|
|
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review confirms that the device met all material, assembly and performance specifications.Visual inspection of the device revealed that the delivery wire was kinked, likely due to handling.The main coil was found to be kinked and appears to be physically detached at the main coil junction.During functional testing, coil could not be advanced through the introducer sheath as the main coil was stuck in the tip of the introducer sheath.The physician may have perceived the observed damages (coil delivery wire kink) as break and reported it as such; however, no break was identified on the device.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
|
| |
|
Event or Problem Description
|
|
It was reported that during a coil embolization procedure, the coil (subject device) experienced resistance during insertion into microcatheter and upon removal, the coil delivery wire broke.The coil was retrieved along with the microcatheter.There were no reported clinical consequences to the patient.
|
| |
|