• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED CORPORATION 4.5MM STERLING LIGHTNING; ARTHROSCOPIC SHAVER/BUR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED CORPORATION 4.5MM STERLING LIGHTNING; ARTHROSCOPIC SHAVER/BUR Back to Search Results
Catalog Number H9133
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Examination of the device item, h9133, found the bur detached from the inner hub, due to oversized inner hub dimension.This is a manufacturing failure and is currently being investigated.A review of the device history review found no anomalies that would contribute to this issue.A two-year review of complaint history revealed there has been a total of 36 complaints, regarding 46 devices, for this device family and failure mode.During the same time frame (b)(4) devices were manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be.(b)(4).Per the instructions for use, the user is advised the following; do not use equipment if upon receipt, package is opened, damaged, or shows signs of tampering.Continually check for overheating, if overheating is sensed immediately discontinue use.Do not use burs for plunge cutting.Injury or damage could occur.Do not use shaver bur if they show any signs of damage.Inspect for bent, dull, or damaged blades and burs before each use.Do not apply excessive loading/force on the shaver blade or bur.Direct contact of the rotating cutting edge of shaver blades and burs with metallic surfaces and/or other hard surfaces such as arthroscopes, cannulas, or other instruments can cause damage to the devices, shaver blade or bur should be examined for damage and replaced if necessary.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The customer reported the "h9133 - 4.5 mm sterling, lightning, "did not work properly and broke off".This was reported with no patient injury or surgical delay.Based on the decision tree results and provided information, this will be reported as a device malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM STERLING LIGHTNING
Common Device Name
ARTHROSCOPIC SHAVER/BUR
Manufacturer (Section D)
CONMED CORPORATION
11311 concept boulevard
largo FL 33773
MDR Report Key7838737
Report Number1017294-2018-00140
Device Sequence Number1175719
Product Code GFF
UDI-Device Identifier10845854029671
UDI-Public(01)10845854029671(17)230308(10)913214
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2023
Device Catalogue NumberH9133
Device Lot Number913214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2018
Initial Report FDA Received Date08/31/2018
Patient Sequence Number1
-
-