| Model Number |
8888119369P |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during dialysis treatment, the second device fell out from the patient.It was reported the securement wing was sutured.And there were nodules on the silver ion sleeve.It was reported the catheter was removed and another one was inserted.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: the device was received covered in blood with fluid dried inside.The bubbles on the silver sleeve are a normal part of the function of the device when the silver has been dissolved.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|