• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR SIZE17 LATERAL REDUCED NECK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. LINEAR HIP; STEM, HIP, FEMORAL, LINEAR SIZE17 LATERAL REDUCED NECK Back to Search Results
Model Number 425-11-017
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 08/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having a peri-prosthetic fracture of the femur.The surgeon removed and replaced the stem, head, sleeve, and liner.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to a peri-prosthetic fracture.The previous surgery and the revision detailed in this investigation occurred 5.2 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a peri-prosthetic fracture.There were no findings during this investigation that indicate that the reported device (s) was the source or had a direct connection with the event.There are many factors that may contribute to the event that are outside the control of djo surgical such as: excessive loads or torques, poor bone quality, patient bone deterioration, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR HIP
Common Device Name
STEM, HIP, FEMORAL, LINEAR SIZE17 LATERAL REDUCED NECK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7839454
Report Number1644408-2018-00824
Device Sequence Number13938135
Product Code LPH
UDI-Device Identifier00888912078214
UDI-Public(01)00888912078214
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K991325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2018
Device Model Number425-11-017
Device Catalogue Number425-11-017
Device Lot Number124C1006
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2018
Supplement Date Received by Manufacturer09/12/2018
Initial Report FDA Received Date08/31/2018
Supplement Report FDA Received Date09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-000, LOT 902B1099; 497-40-000, LOT 641C1030; 932-40-256, LOT 743F1067
Outcome Attributed to Adverse Event Required Intervention;
Patient Age87 YR
-
-