(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The reported patient effect of dissection is listed in the hi-torque guide wires instructions for use as a known patient effect of coronary stenting procedures.The investigation determined a conclusive cause for the reported detachment cannot be determined.The reported detachment possibly contributed to the reported dissection and thrombosis; however a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.(b)(4).(b)(4).
|