| Model Number |
2067L |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: analysis found that the radiofrequency head had an out of specification printed circuit board (pcb) assembly.When the head was connected to the programmer, it would cause a programmer shutdown.The unit also failed testing and was unable to interrogate.The head has been scrapped and replaced.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The radiofrequency (rf) head that was returned after being dropped subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
|
| |
|