• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Premature Discharge of Battery (1057); Moisture Damage (1405)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a power (battery life w/ moisture) issue.The reporter stated there was moisture corrosion in the battery compartment and denied damage to the pump and the battery cap.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Follow-up #1: date of submission: 18-sep-2018.Device evaluation: the device has been returned and evaluated by product analysis on 06-sep-2018 with the following findings: on investigation, the returned battery cap was intact and without damage and able fit properly to the pump.Evidence of moisture was confirmed in the battery compartment.The pump was unable to power on for testing; the device was unresponsive.Due to this failure to power on, the alleged battery life complaint was not able to be investigated.Unrelated to the original complaint, the battery compartment was noted to be cracked during the investigation through which moisture ingress occurred confirmed by leak testing.There was no damage, defect or contamination of the pump¿s interior components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7840562
Report Number2531779-2018-15852
Device Sequence Number1940565
Product Code OYC
UDI-Device Identifier10840406100372
UDI-Public10840406100372
Combination Product (Y/N)Y
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Was the Report Sent to FDA? Yes
Device Age29 MO
Initial Date Received by Manufacturer 08/13/2018
Supplement Date Received by Manufacturer08/13/2018
Initial Report FDA Received Date09/03/2018
Supplement Report FDA Received Date09/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age14 YR
Patient Weight17
-
-