• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US TACIT QUICK ANCHOR #2/0, W/V-5; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US TACIT QUICK ANCHOR #2/0, W/V-5; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 212236
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/10/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Visual observation under magnification confirms that the proximal end of the anchor is stressed, indicating that the anchor was used.No anomalies were observed to the threads of the anchor or the inserter.No technique information was provided that would suggest if the aforementioned causes contributed to this event.Although the complaint can be confirmed, we cannot discern a root cause for this failure.A batch record review has been conducted and our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed no other complaint of any kind for this lot released to distribution.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the (b)(4) complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that during an unspecified plastic surgery, it was observed that the tacit quick anchor #2/0, w/v-5 device was loose and came off from the bone hole after insertion.Another anchor was used to complete the case.There were no adverse consequences to the patient.It was not reported if there was a delay in the surgical procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACIT QUICK ANCHOR #2/0, W/V-5
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7840864
Report Number1221934-2018-54251
Device Sequence Number1271741
Product Code MBI
UDI-Device Identifier10886705002023
UDI-Public10886705002023
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K961094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/10/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number212236
Device Lot Number3754547
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2014
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2014
Initial Report FDA Received Date09/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-