DEPUY MITEK LLC US HEALIX BR ANCHOR, 5.5 MM - W/ORTHOCORD & NEEDLES; SOFT-TISSUE ANCHOR, BIOABSORBABLE
|
Back to Search Results |
|
| Catalog Number |
222232 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/10/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Visual observation confirms the anchor is broken in half and the distal end is broken as well.Anchor breakages are typically associated with off axis insertion, levering during insertion, hard bone quality or using incorrect instrumentation for preparing bone hole.No information was provided on how the hole was prepared.The root cause can be a combination of the above mentioned issues.Furthermore, no lot numbers were provided which precludes conducting a batch history review or a lot specific search in the complaints handling system.At this point in time, no further action is warranted.If additional information is made available to mitek complaints in future, this complaint will be re-opened and re-evaluated.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4)-incomplete.The lot number is currently unavailable.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the affiliate in (b)(6) that during rotator cuff repair surgical procedure, it was observed that the healix br anchor, 5.5 mm - w/orthocord & needles device broke while screwing in.It was reportedly not off-axis, bone structure normal to hard.There was a delay of 15 minutes.It was reported that the device was able to be removed.It was not reported if a spare device was available for use.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|