WARSAW ORTHOPEDICS ATLAS CABLE SYSTEM Back to Search Results
Catalog Number 826-213
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-op diagnosis: axial dislocation with incomplete spinal nerve injury, posterior arch nonunion, lumbar 5/ sacral 1 disc herniation procedure: open reduction of axial dislocation via posterior approach, decompression of spinal canal and cantilever beam internal fixation it was reported that intra-op, a single cable with integral crimp broke during the procedure.The product came in contact with the patient.There are no known complications to the patient in the reported event.
 
Additional Manufacturer Narrative
Product analysis: the cable was returned with a fracture at the weld junction.A microscopic review of the fracture face reveals a ductile failure with a concaved surface consistent with necking near the failure point.This is consistent with a tensile overload on the axis of the cable.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS CABLE SYSTEM
Common Device Name
CABLE
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7841130
Report Number1030489-2018-01200
Device Sequence Number2062231
Product Code ISN
UDI-Device Identifier00673978047980
UDI-Public00673978047980
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K892651
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2026
Device Catalogue Number826-213
Device Lot Number0640587W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer10/22/2018
Initial Report FDA Received Date09/04/2018
Supplement Report FDA Received Date10/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age39 YR
-
-