| Catalog Number |
25740018570 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Occupation: medical equipment company technician/representative although it is unknown if the device contributed to the reported event, we are filing this report for notification purposes.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Pre-operative diagnosis: proximal junctional kyphosis procedure: implant replacement and debridement levels implanted: s2 it was reported that on an unknown date, post-op, after fusion was performed on t10/s2ai due to adult spine deformity, infection occurred (corynebacteria).At the time of debridement, it was confirmed that the set screw backed out.This time the replacement of the reported product was also performed along with the replacement of the rod.Revision surgery was performed as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual and optical inspection confirmed the screw has been used and the screw's lower pivot cap has been damaged.The pivot cap of the screw appears to have been worn down from repetitive contact from a rod that was placed in the saddle of the screw.This is consistent with repeated use or contact.The threads of the screw show no signs of being damaged during entry or removal.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|