| Catalog Number |
9735129 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.Manufacturing date was unavailable at the time of filing.The instrument was returned to the manufacturer for analysis.The clamp had debris in the threads causing difficulty setting the clamp.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the screw to tighten the spine clamp was starting to strip.There was no patient present.
|
| |
|