| Model Number |
MMT-523NAS |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump was rejecting the new batteries and squirt insulin when reservoir was over filled and act button had to press twice.The customer¿s blood glucose level was unknown at the time of the incident.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).All buttons functioning properly.No moisture/damage during visual inspection noted.Device received with all operating currents within specification and passed self-test, unexpected restart error test and displacement test.No unexpected low battery, weak battery, battery out limit or failed battery test alarms noted during testing.Device received compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform basic occlusion test, occlusion test, prime test, excessive no delivery test or verify no delivery alarm or prime/fill anomaly due to compromised force sensor system alarm.However, device passed rewind test and displacement test.Device had broken battery tube threads.
|
| |
|
Search Alerts/Recalls
|