• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS TG GPO-PAP; TRIGLYCERIDES GPO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS TG GPO-PAP; TRIGLYCERIDES GPO Back to Search Results
Catalog Number 04657594190
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred in (b)(6).
 
Event or Problem Description
The customer complained of discrepant results for 1 patient sample tested for trigl triglycerides (trigl) on a cobas c 111.The initial trigl result was 9.50 mmol/l with a data flag.The sample was diluted 1:10 and repeated with a result of 3.01 mmol/l.The sample was repeated neat and the result was 8.97 mmol/l with no data flag.The result of 8.97 mmol/l was reported outside of the laboratory.The sample was very lipemic so the high result was believed by the technician.On (b)(6) 2018 a new sample was obtained from the patient and the trigl result was 0.76 mmol/l.Based on the lower result, the original sample was repeated and the result was 3.23 mmol/l and the sample looked less lipemic.The patient was fasting when both samples were obtained.The customer is confident there was no sample mismatch.There was no allegation that an adverse event occurred.The c 111 serial number was (b)(4).
 
Additional Manufacturer Narrative
The patient sample was submitted for investigation and tested on a c 111 with reagent lot 312720.The investigation received undiluted results of 3.30 mmol/l with no data flag and 3.42 mmol/l with no data flag.Qc results at the investigation site were within range.The customer's complaint could not be reproduced.The investigation was unable to find a definitive root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TG GPO-PAP
Common Device Name
TRIGLYCERIDES GPO
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7841865
Report Number1823260-2018-02947
Device Sequence Number1940444
Product Code CDT
UDI-Device Identifier04015630921959
UDI-Public4015630921959
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K873049
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04657594190
Device Lot Number31272001
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer08/14/2018
Initial Report FDA Received Date09/04/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTERIA-VITA; CO-DAFALGAN; IRFEN; PANTAPRAZOLE; SEQUASE; VALACICLOVIR; ZOVIRAX CREME