• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems No Device Output (1435); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/09/2018
Event Type  malfunction  
Manufacturer Narrative
Patient identifier not available from the site.No parts have been returned to the manufacturer for analysis.Device manufacturing date is unavailable.
 
Event Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the surgeon had difficulty registering the patient.Site utilized a scan for an infant 5 months prior to case and surgeon was unable to complete a good registration.Surgeon opted to abort navigation because the system was only picking up one side of the patient¿s anatomy.There was a reported delay to procedure of 30 minutes due to this issue.There was no impact to patient.
 
Manufacturer Narrative
Add: patient identifier.Correction: initial reporter.
 
Manufacturer Narrative
A software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Follow-up with the rep determined that the issue has not occurred since the initial incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
erika mitchellette
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7842141
MDR Text Key119156606
Report Number1723170-2018-04366
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age6 MO
Patient Weight7
-
-