Model Number 9735665 |
Device Problems
No Device Output (1435); Communication or Transmission Problem (2896)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient identifier not available from the site.No parts have been returned to the manufacturer for analysis.Device manufacturing date is unavailable.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the surgeon had difficulty registering the patient.Site utilized a scan for an infant 5 months prior to case and surgeon was unable to complete a good registration.Surgeon opted to abort navigation because the system was only picking up one side of the patient¿s anatomy.There was a reported delay to procedure of 30 minutes due to this issue.There was no impact to patient.
|
|
Manufacturer Narrative
|
Add: patient identifier.Correction: initial reporter.
|
|
Manufacturer Narrative
|
A software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Follow-up with the rep determined that the issue has not occurred since the initial incident.
|
|