Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 135, 152, 196, 138 and 203 mg/dl.The customer's expected fasting blood glucose test result range is 103 - 118mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the bathroom.The test strip lot manufacturer's expiration date is 09/19/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: (b)(6).
|