A review of the device history records was performed for the indicated packaging and assembly component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The july 2018 angiodynamics complaint report was reviewed for the angiovac product family and the failure mode "patient injury/death." no adverse trend was indicated.The angiovac cannula was not returned for evaluation as there was no report of device malfunction.As stated by the physician, the cause of death was believe to have been a "hemorrhage related to ivc and anticoagulation." the directions for use provided with the angiovac device contain the following statements: "warnings- selection of the patient as a candidate for use with this device and for such procedures as it is intended is the physicians' sole responsibility.The outcome is dependent on many variables including, patient pathology, surgical procedure, and perfusion procedure/technique.The benefits of use of this device must be weighed against the risks including risks of systemic anticoagulation and must be assessed by the prescribing physician.As with all medical devices, this device and ancillary equipment are to be used by trained physicians only.Specifically, this device is to be used only by medical personnel experienced with initiating and monitoring extracorporeal bypass and by physicians trained and experienced using surgical and/or percutaneous (seldinger) vascular access techniques.Adverse events- this device, as do all extracorporeal blood vessel devices, has possible side effects, which include but are not limited to infections, blood loss, thrombus formation, embolic events, vessel, ventricular or valvular damage and complications of percutaneous or surgical insertion techniques.These may occur if the instructions for use are not followed.Possible complications include those normally associated with large bore surgical and/or percutaneous vessel catheterization/cannulation, anticoagulation, extracorporeal circulation and application of intravascular introducer systems which include but are not limited to: death pulmonary embolism".(b)(4).
|