• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN AUTOSONIX; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN AUTOSONIX; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 012001
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during thorascopic lobectomy, the device was used but it could not cut.A new device was used to continue.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one device.The jaw operated properly upon actuation of the handle but failed a grasping test.The device was disassembled and damage to the probe housing grove was observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the damage to the probe housing groove is consistent with excessive material or an obstruction being compressed in the jaws as the trigger is connected to the groove and the jaws are closed.Any damage to the groove interface will result in the handle bottoming out and the jaws not closing completely.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSONIX
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7843096
Report Number1219930-2018-04828
Device Sequence Number16527912
Product Code LFL
UDI-Device Identifier10884521058163
UDI-Public10884521058163
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Model Number012001
Device Catalogue Number012001
Device Lot NumberN4F1472X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Initial Date Received by Manufacturer 08/16/2018
Supplement Date Received by Manufacturer10/31/2018
Initial Report FDA Received Date09/04/2018
Supplement Report FDA Received Date11/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-