Model Number S7
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Date 08/09/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A software analysis was performed using the patient exams.The manufacturer representative noticed that the spacing in the slices were different when low and high dosages meet.The manufacturer representative suggested that the site uses the same spacing throughout even when changing to a lower dosage.
 
Event Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the images loaded were missing anatomy when viewed in the dicom.Loading on the cranial application results in images missing from the top of the head.There was no patient present when this issue was observed.
 
Manufacturer Narrative
A software investigation analysis was initiated to determine the probable cause of the issue through review of the reported issue.Analysis found that the behavior described is the intended behavior of the software.The exams were not to protocol, with non-contiguous slices.Software is functioning as designed.Analysis found that the software functioned as designed.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
parker desautel
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7843330
MDR Text Key119106887
Report Number1723170-2018-04363
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-