Catalog Number 1010480-HJ
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.It is possible the unspecified balloon dilatation catheter met resistance during advancement over the guide wire due to a buildup of procedural contaminants (blood, contrast) thus resulting in the reported difficult to position and difficult to remove; however this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a 95% stenosed and de novo lesion in the mid right coronary artery.A 014 hi-torque pilot guide wire was advanced to the lesion and a 2.0x15mm unspecified balloon dilatation catheter (bdc) was advanced; however, resistance was met and the bdc became stuck on the guide wire and could not be removed.The guide wire and bdc were removed as a single unit.The lesion was pre-dilated and the procedure was successfully completed with the deployment of a 3.0x32mm unspecified drug eluting stent.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE PILOT GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7843442
Report Number2024168-2018-06852
Device Sequence Number2062206
Product Code DQX
UDI-Device Identifier08717648073861
UDI-Public08717648073861
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K060449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Catalogue Number1010480-HJ
Device Lot Number8030171
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2018
Initial Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-