WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.4MM VA-LCP 2-CLMN VLR DSTL RADIUS PL 6H HD/5H SHAFT/RIGHT; PLATE,FIXATION,BONE
|
Back to Search Results |
|
| Model Number |
02.111.650 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Not available for reporting.No information available.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is synthes sales consultant.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that, on (b)(6) 2018, a broken distal radius plate was explanted.The original implant date is unknown.It is unknown if there was any surgical delay.The patient and surgical outcome are unknown.This is report 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
The plate was originally implanted on (b)(6) 2018 to to correct a distal radius fracture caused by motor vehicle accident.The plate broke post-operatively.The issue was discovered during a post-op follow-up visit.It is unknown if patient was symptomatic.Revision surgery was uneventful.There was no time delay and patient outcome was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient height reported as 4 ft., 11" in.Additional information provided.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: part: 02.111.650; lot: 9432503; manufacturing site: mezzovico; release to warehouse date: (b)(6)2015.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.This implant was manufactured from forging blank 60062491 lot 9397619 which was manufactured in mezzovico as well.The raw material certificate 18039 was reviewed and the used material was according to iso-5832-1 specification for implants for surgery.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.The following investigation flow was performed: broken: visual inspection performed at customer quality (cq) confirmed the broken condition on the returned volar plate which agrees with the reported complaint condition.The plate is broken transversely at about half of its length and both broken portions were returned.No new issues were identified.Dimensional analysis and document/specification review: review of drawings from the date of manufacture to current showed no design issues: -va-lcp volar plate drawing dimensional inspection performed measured the thickness at the broken site as 2.01 mm(ca-814) that falls within the specification of 2mm +/- 0.1 mm and tolerance specified.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing process.Furthermore, the raw material certificate reviewed showed the used material was according to iso-5832-1 specification for implants for surgery.Also, the radial plate design and clinical risk management (dcrm) document, reviewed during investigation was found to adequately address the complaint condition.A definitive root cause could not be determined with the provided information.This complaint is confirmed, however no product design issues or manufacturing discrepancies were identified during this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|