Evaluated 8 open/used reservoirs.Inspected reservoirs, snap-cap, and septum for anomalies, none were found.Ran occlusion test as follows: reservoir filled with diluent and connected to a new infusion set.Installed reservoirs & connector into a paradigm pump, performed pump manual prime, saw fluid flow through infusion set and out catheter tip.Conclusion: reservoirs not occluded.Also, inspected 6 of 8 reservoir¿s snap-caps width dimension per (b)(4).Also, using a new lab infusion sets connect found reservoir's snap-caps too loose to connect/lock.2 remaining's checked reservoirs using a new lab infusion set.Found reservoirs easy to connected/release.No connector anomalies were observed during analysis.(b)(4).
|