Model Number 0603870
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during port catheter insertion, there was alleged swelling and flexing of the tunneling mandrel.It was further reported that the tunneler was unable to be placed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the lot met all release criteria.Manufacturing records were reviewed (dhr, mrr¿s, scrap and mfg.Process changes) and there was no evidence found that the failure mode reported in this complaint was caused by the mfg.Process.Investigation summary: the devices returned were a 7 fr macrointroducer with both dilator and sheath.Visual, microscopic, and tactile evaluations were performed.The complaint of tunneler (introducer) damage is confirmed and found specifically to be damage to the dilator tip.The introducer manifested mechanical damage to the tip of the dilator such that the distal orifice was no longer circular.The root cause cannot be determined based upon available information.It is unknown if procedural issues contributed to the reported event; the damage found on the dilator tip and the report of difficulty of insertion are consistent with resistance during insertion, such as from poor insertion technique.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during port catheter insertion, there was alleged swelling and flexing of the tunneling mandrel.It was further reported that the tunneler was unable to be placed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRI L/P PORT W/GROS 7.0F
Common Device Name
IMPLANTABLE PORT
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7844436
Report Number3006260740-2018-02344
Device Sequence Number11250421
Product Code LJT
UDI-Device Identifier00801741025839
UDI-Public(01)00801741025839
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K924250
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0603870
Device Catalogue Number0603870
Device Lot NumberREBX1612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Initial Date Received by Manufacturer 08/09/2018
Supplement Date Received by Manufacturer11/19/2018
Initial Report FDA Received Date09/04/2018
Supplement Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR