Model Number EX061003CS
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the vascular stent allegedly jumped approximately 10cm forward as it was being deployed in the superficial femoral artery (sfa) through an antegrade approach.It was further reported that an additional 100mm vascular stent was used however, the stent also jumped and twisted during deployment, misplacing this vascular stent further.Therefore, a balloon expandable stent was then used to straighten the 100mm vascular stent, bridged the gap, covered the lesion, and completed the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT SOLO VASCULAR STENT SYSTEM
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key7844665
Report Number9681442-2018-00159
Device Sequence Number1129833
Product Code NIP
UDI-Device Identifier04049519001272
UDI-Public(01)04049519001272
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEX061003CS
Device Catalogue NumberEX062003CL
Device Lot NumberANCN2479
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2018
Initial Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
-
-