As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported that the vascular stent allegedly jumped approximately 10cm forward as it was being deployed in the superficial femoral artery (sfa) through an antegrade approach.It was further reported that an additional 100mm vascular stent was used however, the stent also jumped and twisted during deployment, misplacing this vascular stent further.Therefore, a balloon expandable stent was then used to straighten the 100mm vascular stent, bridged the gap, covered the lesion, and completed the procedure.There was no reported patient injury.
|