| Catalog Number |
07026889190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer complained of (b)(6) elecsys anti-hcv ii immunoassay results for 1 patient sample on a cobas 8000 e 602 module and a cobas 8000 e 801 module.The e 602 module serial number was (b)(6) and the e 801 module serial number was (b)(6).The initial anti-hcv ii result was (b)(6) on the e 602 module with reagent lot number 31063600.The patient questioned the result.On (b)(6) 2018, the repeat anti-hcv ii result was (b)(6) on the e 801 module with reagent lot number 31072500.The customer then sent the sample to a different laboratory and the anti-hcv result was (b)(6) with the immunoblot method.The specific date of the immunoblot result was not provided.There were erroneous results reported outside of the laboratory.There was no allegation of an adverse event.This medwatch will cover anti-hcv ii reagent lot number 31063600.Refer to medwatch with patient identifier (b)(6) for information on anti-hcv ii reagent lot number 31072500.
|
| |
|
Additional Manufacturer Narrative
|
|
Sample from the patient was submitted for investigation and further testing by the fujirebio innolia hcv score confirmed the negative result.The negative elecsys anti-hcv ii immunoassay result was therefore believed to be correct.
|
| |
|